X-RAY COMPLIANCE TEST METER Model # TBM-XR w/Built-in Detector Model # TBM-EXR w/ Separable Probe FEATURES: * X-RAY MACHINES, TV, and MONITORS * LUGGAGE and PEOPLE SCANNERS * ONE HAND OPERATION - Both Models * DIGITAL READOUT: 6 digit-rate, 6 digits integrate * SMALL LIGHTWEIGHT: 2-3 lbs * SEES AXIALLY BELOW 15 KeV GAMMA OR X-RAYS * FLAT RATE RESPONSE - AIR EQUIVALENT * SCINTILLATOR - 15 KeV - 500 KeV WIDE RANGE APPLICATION: Model TBM-XR and TBM-EXR allow rapid, accurate compliance testing of a wide variety of X-Ray machines, TV's and computer monitors, luggage and people x-ray scanners. Protects the health and safety of airport and shipping personnel while they protect the public. The TBM-XR and TBM-EXR are now smaller and lighter. Based on newest, more stable, essentially drift-free electrometer technology. DESCRIPTION: The TBM-XR and TBM-EXR consist of an air-equivalent organic scintillation probe coupled to a stable solid state MOSFET input electrometer with built in A to D converter to read out directly in mR/h or mR. Rate Range is 0.01 mR/h to 100.00 mR/h in one range. Dose range is 0.01 mR to 10 R. TBM-EXR SPECIFICATIONS: Detector: Air equivalent plastic scintillator. Window: Light-tight 0.9 mg/cm$^{2}$ Mylar laminate. 10 sq cm area. 36 mm (1.4") dia. Readout: LCD 8 digits Indicator Light: Green LED 10 pulses/min per mR/h. Red LED Over-range Indicator. Range: Rate 6 digits 0.1 mR/h to 999.9 R/h in one range. Integrate 6 digits 0.01 mR to 99.9R. Photon Energy Range: 15 KeV - 500 KeV ± 20% High Voltage: Fully user adjustable (internal pot). X-Ray Rep Rate: Use either rate or integrate mode for pulses with repetition rate of less than 1.5 sec. Use Integrate Mode for greater than 1.5 sec. T$_{\textrm{A}}$ TECHNICAL ASSOCIATES ------------------------------------------------------------------------------------------ 7051 ETON AVENUE * CANOGA PARK, CA 91303 TELEPHONE (818) 883-7043 * FAX(818) 883-6103 $Revision: 1.2 $ X-RAY COMPLIANCE TEST METER Model # TBM-XR w/Built-in Detector Model # TBM-EXR w/ Separable Probe SPECIFICATIONS CONTINUED: Electrometer: Solid State MOSFET input. Electronics: Adj H. V. Supply, Pulse Shaper. A-D converter LCD drivers. Test and display functions. Pulse Duration and Content: With sufficient repetitions, TBM-XR and TBM-EXR can accurately integrate pulses of less than a microsecond duration and longer and including continuous emission. Pulse content of 0.5 nanoR are measurable with sufficient repetition rate to give at least .01 mR/hr for at least 5 to 10 seconds. Environment: Temperature of Operation from -20$^{\textrm{o}}$ to 50$^{\textrm{o}}$ C Relative Humidity 0 - 95% Radio-frequency 10 mW/cm$^{2}$ up to 4 GHz affects meter viewing < ± 10% Magnetic 200 Gauss ± 5% effect Geotropism < ± 2% effect Response Time: 2-3 seconds for rate and for integrate. Time to Zero: If count-rate goes to zero, the digital display will hold the old count for 12 seconds before displaying "0.0." Batteries: Front panel battery test is provided. 6 ea. (AA) NEDA 15A - over 100 hrs of operation. Dimensions: TBM-XR (excluding handle): 5.2"H x 4.2"W x 7.3"L TBM-EXR (excluding handle, probe, & probe holder): 3.2"H x 4.2"W x 7.3"L Weight: TBM-XR complete with batteries and internal detector: 2.4 lbs. TBM-EXR complete with batteries, external detector and probe holder: 3.0 lbs. Options: Readout in Si units nGy and nGy/h. Ra-UBG - Check source. Other entrance window material or thickness. High voltage test button. 5 cm nylon positioning bumpers. T$_{\textrm{A}}$ TECHNICAL ASSOCIATES ------------------------------------------------------------------------------------------ 7051 ETON AVENUE * CANOGA PARK, CA 91303 TELEPHONE (818) 883-7043 * FAX(818) 883-6103 $Revision: 1.2 $