SCINTILLATION WELL COUNTING SYSTEM Model # SSS-11([fig] ([fig] SSS-11) FEATURES: * COMPLETELY SHIELDED WELL COUNTER * UNIVERSALLY USEFUL WITH STANDARD SCALERS, RATEMETERS, AND SPECTROMETERS * ACCEPTS BEAKERS, FLASKS, PLANCHETS, ETC., IN ADDITION TO TEST TUBES DESCRIPTION: This unit is a stable but extremely sensitive detector system created for universal use in research, clinical, and industrial laboratories. It is compatible with all standard nuclear electronic equipment designed for scintillation counting. SPECIFICATIONS: Sample: Test tubes up to 21/32" dia. in well; or beakers, flasks, planchets, etc. up to 2-1/2" dia. atop crystal. Shielding: 2" lead minimum in all directions. Power Requirements: HV 800-1300 V regulated. Connectors: HV and signal output MHV low noise teflon insulated. Also may be matched to standard electronic equipment. BNC connector Optional. Includes: MODEL LS-21 Shield. MODEL PGS-3LW Well Crystal Scintillation Probe. SSS-11 Schematic T_{\textrm{A}} TECHNICAL ASSOCIATES ------------------------------------------------------------------------------------------ 7051 ETON AVENUE * CANOGA PARK, CA 91303 TELEPHONE (818) 883-7043 * FAX(818) 883-6103 $Revision: 1.1 $